Herzog (PAC) |
HCP 852 Automated Pour Point and Cloud Point Analyzer |
 | The HCP 852 Automated Pour Point and Cloud Point Analyzer removes human error from the equation. The only automatic analyzer that performs measurements exactly according to their respective ASTM test procedures, it controls each step with microprocessor precision. Sophisticated robotics control all sample movement. And to complete the process, a high resolution CCD video detection system simultaneously examines the entire sample for movement (pour point) and crystal formation (cloud point). |
|
Standards |
ASTM D 97, D 2500, D 6892; IP 15, IP 219, ISO 219, ISO 3015, DIN ISO 3016 and DIN EN 24015 |
|
Features |
Automates ASTM D 97, D 2500, D 6892; IP 15, IP 219, ISO 219, ISO 3015, DIN ISO 3016 and DIN EN 24015 testing with microprocessor precision
Provides unparalleled precision and accuracy
Stand-alone or networked operation
Simple to operate and maintain |
|
Applications |
|
|
Tech. parameters |
Viscosity
range: |
0.3
to 1000 mPa.s (extended measuring range upon request)
Repeatability: < 0.35 %
Reproducibility: < 0.7 % |
Time
measuring range: |
0 to 250 s (extended measuring range upon request)
Resolution: +/- 0.01 s
Accuracy: < 0.02 s |
Temperature
range: |
+10
to 70 °C (extended measuring range upon request)
Resolution: +/- 0.01 °C
Accuracy: < 0.05 °C |
Capillary
diameters available: |
1.6; 1.8; 3; 4 mm (extended measuring range upon
request)
material: glass |
Ball
diameters available: |
1.5;
2.5; 3 mm (extended measuring range upon request)
material: stainless steel |
Power: |
85 to 264 V AC, 50 .... 60 Hz, 75 VA |
Dimensions: |
(W x H
x D) 270 x 340 x 310 mm |
Weight: |
15 kg (150 N) |
Ambient
temperature: |
15 to
35 °C |
Humidity: |
80 % RH, 35 °C and below |
Accessories: |
Printer
and external keyboard recommended |
|
|
Documentation |
|
|
|